Applications

  • Environmental noise assessements
  • Building Acoustic
  • Room acoustics - STIPA
  • Noise monitoring
  • Product noise testen
  • Vibration measurements
  • Noise in the workplace
  • Infrasound
  • Ultrasound

Highlights

  • Easy connectivity via built in WLAN and 3G/4G LTE modem.
  • Large colour touchscreen
  • Intuitive user interface and custom-made user setups.
  • Easy to use.

About

text

Features

text
  • Precision sound level meter and frequency analyser according to class 1.
  • Easy connectivity via built in WLAN and 3G/4G. LTE modem (Nor145).
  • Large colour touchscreen (4.3”).
  • Real push keys for quick operation in challenging environments.
  • Intuitive user interface with graphical icons for selection of measurement mode and custom-made user setups.
  • Voice, text notes and built-in GPS for documentation of the measurements.
  • Wide frequency range (0,4 Hz – 40 kHz in 1/3 octave band and FFT).
  • Parallel 1/3 octaves and FFT analysis.
  • 120 dB measurement range.

Standarder

text
  • ISO 717-1, ISO 717-2, ISO 10140-2, ISO 10140-3, ISO 10140-4, ISO 16283-1, ISO 16283-2, ISO 16283-3, ISO 354, ISO 3382-2, ISO 10052, ISO 1996-1, ISO 1996-2
  • IEC 61260-1, IEC 61260-2, IEC 61260-3, IEC 61672-1, IEC 61672-2, IEC 61672-3
  • EN 61010-1
  • ASTM E1007-21, ASTM C423-23, ASTM E336-23, ASTM E413-22, ASTM E90-9
  • ANSI S1.4, ANSI S1.11
  • DIN 45657, DIN 4109-4, TA-Lärm
  • NS 4814, NS 4815, NS 8172
  • BS EN ISO 16283-1, BS EN ISO 16283-2, BS EN ISO 16283-3
  • SS 25267
  • SIA 181

Applications

text

Selecting the different application is easy using the Application menu appearing after power on.
You may also create own favorite setups indicated as smaller icons in the startup picture.

Environmental

text

Nor145 is designed to be the preferred tool for all types of environmental noise measurements.

Measurements with markers, audio recordings and event triggered pictures are easily made and synced to your project via the NorRemote app for your phone or tablet (android). The large 4.3” display gives you all the necessary information. Up to 60 measurement parameters may be logged simultaneously.

The advanced trigger system enables different trigger levels for Day, Evening or Night with event triggers to collect markers, audio recordings and pictures. The data can be displayed remotely in a web browser on any device, or by the NorRemote app. From the same remote device, you may change settings or simply check the status of your Nor145.

Nor145 is also optimised for use with Norsonic NorCloud, Norsonics’ cloud service for noise measurement, control and reporting, all easy connected over the internal 4G modem.

Features

  • Twin time profiles with resolution from 5 ms and additional Moving report with trigger possibility.
  • Extensive trigger system for reports, audio recording, camera and digital output lines.
  • Voice, text and picture notes.
  • 5 independent event triggers (LDEN support)
  • Automatic detection of impulsive noise (ISO/PAS 1996-3:2022).
  • Advanced marker management.
  • Full remote-control support via NorVirtual app for smartphones.
  • Seamless connection to NorCloud for unattended monitoring and reporting.
  • 0-20 sec graphical back erase / pause function.
  • 0-120 sec Audio pre-trigger.
  • Seamless integration to post-processing programs and Excel.
  • API interface for system integrators.

General analyser (environmental) mode has a selection of different display views. Portrait and Landscape. All Portrait views can either be a single or dual type frame.

Available views in Environmental / General mode are:

Nor145 screenshot

Sound level meter display

 

 

 

 

 

 

 

 

 

 

FFT – Displays the FFT spectra

FFT – Displays the FFT spectra

FFT (landscape) Displays the FFT spectra as landscape

FFT (landscape) Displays the FFT spectra as landscape

 

 

 

 

 

 

 

 

 

 

Level vs frequency (L/f) display single view – portrait

Level vs frequency (L/f) display single view – portrait

Level vs frequency (L/f) display single view – landscape

Level vs frequency (L/f) display single view – landscape

 

 

 

 

 

 

 

 

 

 

 

Level vs time (L/t) display single view – portrait

Level vs time (L/t) display single view – portrait

Level vs time (L/t) display single view – landscape

Level vs time (L/t) display single view – landscape

 

 

 

 

 

 

 

 

 

 

A numerical table is associated with all the graphical views – just press the TAB button to toggle between graphical and numerical view. Works both with single and split screen.

Cumulative and Probability Distribution Function /PDF) – landscape view

Cumulative and Probability Distribution Function /PDF) – landscape view

Cumulative and Probability Distribution Function /PDF) – portrait view

Cumulative and Probability Distribution Function /PDF) – portrait view

 

 

 

 

 

 

 

 

 

 

Split view with FFT and 1/3 octave. The link cursor feature connects the cursor in FFT and 1/3 octave displays, making it easy to manoeuvre in the frequency domain

Dual view with L/t and L/f – split view

Dual view with L/t and L/f – split view

 

 

 

 

 

 

 

 

 

 

 

 

 

FFT – Displays the FFT spectra

FFT – Displays the FFT spectra

FFT (landscape) Displays the FFT spectra as landscape

FFT (landscape) Displays the FFT spectra as landscape

 

 

 

 

 

 

 

 

 

 

Graphical back erase in Pause mode. Move cursor to remove unwanted noise events and push continue to resume measurement.

 

 

 

 

 

 

 

 

 

 

 

 

Building acoustics

text

Norsonic is known for its long-time tradition of creating the state-of-the art building acoustic analysers. The Nor145 sound level meter is no exception!

It can be used as a stand-alone single channel building acoustic analyser, or as a remotely controlled building acoustic frontend (wireless) for the advanced Nor850 multichannel software thru its hotspot feature.

The Nor145 offers a built-in signal generator for excitation of the source room level measurements or for excitation of the reverberation time measurements. The results are measured in accordance with the ISO 16283 Standard requirements. With additional background level measurement results, the Nor145 offers a on-board calculation of the final airborne sound insulation indices DnT and Rw in accordance with ISO 717. Of course, the similar possibility is available for impact sound insulation index Ln,w using a tapping machine such as the Nor277.

The reverberation time excitation may alternatively be based on an impulsive source. In any case, the Nor145 calculates the results for T15, T20 and T30 in parallel.

The Nor145 analyser is seamless integrated with the Nor850 software; either as a remote frontend to the Nor850 Measurement System, or for easy post processing in the Nor850 Reporting System.

Features

  • Complete calculation of airborne, façade and impact sound insulation indices in accordance with international and national standards.
  • Wireless single or dual channel measurements using one or two Nor145 controlled from Nor850.
  • Wireless control of the battery-operated power amplifier Nor282.
  • Reverberation time measurements with parallel calculation of T15, T20, T30, Tmax and EDT.
  • Ensemble averaging of reverberation decays.
  • Backward integration of reverberation decays based on impulse excitation.
  • User adjustment of individual RT decay lines.
  • Signal generator with white, pink or bandpass filtered noise.
  • Supports multiple microphone and loudspeaker positions with corresponding on-board energetic or arithmetic averaging.
  • Project overview with information about all individual measurement details and project progressSeamless integration with Nor850 reporting software.




 

Building acoustics mode offers application tailored displays to secure a good work flow and control in your project.

Building Acoustics- Upper frame shows Receiving room level with background noise. Running L/t in lower frame to evaluate if unwanted noise appears in the measurement.

Building acoustic – rating view 1, calculated sound insulation together with the single number index.

Building acoustic – rating view 2, numerical table to present the single number index along with the spectrum adaption terms.

Building acoustic – rating view 3, all numerical results of the analysis plus a level overview graph.

Building acoustic – rating view 4, background noise vs. receiving room levels.

 

 

 

 

 

 

 

 

 

 

 

 

Building acoustic – rating view, The # softkey toggles through specific source positions.

Building acoustic – rating view, R’w rating.

 

 

 

 

 

 

 

 

 

 

 

Building acoustic – RT view.

Two sample pictures out of several views offered for RT measurements.

Building acoustic – manual adaption of RT time is possible.

 

 

 

 

 

 

STI calculations based on the STIPA method

text
  • The STI mode supports the latest revision of the STIPA standard, IEC 60268-16 revision 5.
  • Additional features are background noise correction and auditory correction.
  • Background noise correction can be turned on/off.
  • Auditory correction gives a lower score on the STI value if the STIPA noise signal is too loud or too low. It shall normally be turned on, but the latest revision of the standard recommends it to be possible to turn it off in some special circumstances.
  • The built-in signal generator generates the STIPA signal.
  • A complete measurement sequence takes only 21 seconds.

STIPA mode is easy and intuitive to use. The display is composed of a fixed view with two frames. The upper shows each frequency band, while the lower shows the measured values including a bar graph.

STIPA – By selecting the lower part of the display and pressing the TBL button, the two modulation indexes for each of the STIPA-octaves will be displayed.

Options

text

The Nor145 Analyser is designed to be expanded and upgradeable to give you a complete measurement tool for years to come. Norsonic’s retrofit policy ensures regular software updates with new features and new options followed by a 3 years warranty.

Nor145 basic unit:
includes A, C and Z weighting networks with measurement of Leq, LeqI, LE – sound exposure level, SPL, Lmax, Lmin with parallel detection of Fast, Slow and Impulse time constants, statistical calculation of Ln, and TMAX5. The measurement data is analysed as global values in addition to two parallel time profiles with an adjustable resolution ranging from 5 ms to 24 h. Voice and text annotation included.

Option packages:

Nor145 Advanced Level option kit (e.g. for ENV) containing option: 3, 4, 11, 12.

Nor145 Building Acoustic option kit incl. option 3, 7, 8, 9.

Nor145 Consultant package option kit containing option 3, 4, 7, 8, 11, 12, 13.

Option 1: GPS and support for external IP camera

text

The GPS is used for position identification, and support for exact time synchronisation. Time synchronisation is especially useful when several instruments are used for capturing the same noise event, such as blast monitoring.

With an external IP camera, you can make event triggered pictures. Pictures captured during a measurement is marked as events in the L/t graph and can be view on the Nor145, in NorCloud or in NorReview.

Thttps://norsonic-en.nyg.dev/products/environmental-monitoring/norcloud/he GPS position is stored as a part of the measurement file. It may also be set to be a part of the file name.

This option is not available as retrofit and must therefore be ordered along with the instrument.

Option 3: 1/1 and 1/3 octave band filters

text

1/1 and 1/3 octave filters (0,4 – 20 kHz in 1/3 octave band) including multi-spectre functionality in time profile A.

A 3-order analogue high pass filter with a -3 dB point at 3 Hz may be selected to avoid overloading the input circuitry from low frequency noise.

Option 4: Audio recording and markers

text

This option adds audio recording and marker management to the Nor145. On a Nor145 it is possible to playback and listen to the audio recording via headphones.

Option 4 also adds a sophisticated but easy to use marker management, indispensable for every environmental noise assessment task. The user can select up to ten markers, add names, colour and associate the marker to an action. The marker can either be a single or a toggle marker. The action can be; enable a reference tone, start a recording, take a picture or set one of the digital outputs found on the 15pin socket. It is easy to add markers on an ongoing measurement. Voice and text notes may in the same way as markers be added to an ongoing measurement. The markers can be moved or deleted. Markers may also be added as post processing on the instrument.

The Nor145 offers six system markers that are automatically inserted in the Level versus time profile. These are; Measurement Pause marker, Measurement continue marker, Audio recording marker, Picture marker (from IP camera), Signal overload marker, Battery low marker, Event trigger marker.

The Nor145 is seamless integrated to the PC program NorReview to further evaluate and calculate the noise data and to create customised reports.

Features:

  • 12 or 48 Khz audio recording sampling rate
  • 8, 16 or 24 bit audio recording resolution
  • 0 to 60 dB digital audio recording gain
  • 0 to 120 sec audio recording pre-trigger
  • Innovative marker management system

Option 7: Signal generator

text

Signal generator with sine, white, pink, band passed filtered noise.

Option 8: Reverberation time decay and calculation of T20 and T30

text

Reverberation Time decay and calculation of T20 and T30 based on impulse or noise excitation.

Requires option 3. And option 7 if RT shall be based on noise excitation.

Option 9: Complete Building Acoustic mode

text

Complete Building Acoustic mode with microphone position room averaging in accordance with ISO 16283 as well as sound insulation indices calculated in accordance with ISO 717/1 and /2.

Requires option 3, 7, 8.

Option 11: Enhanced noise assessment package

text

Enhanced noise assessment package including internal web server for remote control via smartphones, pads and PC and additional four triggers for independent setting of different trigger levels during a day.

Requires option 4.

Option 12: NorCloud

text

With this option you are able to connect the instrument to Norsonic’s cloud service; NorCloud.

NorCloud is a Norsonic system to collect, display and process sound level data.

When used together with Nor145 sound analyser, you can measure, analyse and report directly from your web browser. Create projects, assign sound level meters to projects and configure settings such as audio recording, SMS/e-mail alarms and automatic reporting.

Norsonic stores your data safely and conveniently, but you are of course free to download your measurements.

Option 13: FFT

text

FFT sampling rate is 48 kHz with more than 16000 lines. Upper frequency is 22 kHz. Dynamic zoom from 1 – 64. The FFT calculation and 1/3 octave spectrum works in parallel.

Split view with FFT and 1/3 octave. The link cursor feature connects the cursor in FFT and 1/3 octave displays, making it easy to manoeuvre in the frequency domain

Option 14: Ultrasound

text

Ultrasound option including 1/3 octave filters to 40 KHz, 1/1 octave to 31.5 KHz and AU network.

Requires option 5.

Option 16: Build in WLAN and LTE modem (Edge – 4 G)

text

This option is not available as retrofit and must therefore be ordered along with the instrument.

Option 17: STI calculation based on the STIPA method

text
  • The STI mode supports the latest revision of the STIPA standard, IEC 60268-16 revision 5.
  • Additional features are background noise correction and auditory correction.
  • Background noise correction can be turned on/off.
  • Auditory correction gives a lower score on the STI value if the STIPA noise signal is too loud or too low. It shall normally be turned on, but the latest revision of the standard recommends it to be possible to turn it off in some special circumstances.
  • The built-in signal generator generates the STIPA signal.
  • A complete measurement sequence takes only 21 seconds.

STIPA mode is easy and intuitive to use. The display is composed of a fixed view with two frames. The upper shows each frequency band, while the lower shows the measured values including a bar graph.

STIPA – By selecting the lower part of the display and pressing the TBL button, the two modulation indexes for each of the STIPA-octaves will be displayed.

Specifications

text

The Nor145 sound level meter is supplied with ½” preamplifier Nor1209 and the ½” measurement microphone Nor1227.

It conforms to the latest revision of the following National and International standard including amendments, all type 1 / class 1; IEC 61672, IEC 61260, DIN 45657, ANSI S1.4, ANSIS1.11, and ANSI S1.43. It also conforms to the obsolete sound level meter standards IEC 60651 and IEC 60804.

Measured Parameters

Simultaneous measurement of SPL, Leq, LeqI, LMax, LMin, LE, LEI, LPeak, Ln and Tmax5.
Time weighting functions: Fast, Slow and Impulse (Simultaneous).
Spectral weighting functions: A, C and Z – weighting (Simultaneous).
Frequency analysis: 1/1 and 1/3 octave real time filters from 0,4 Hz to 20 kHz.
Statistical calculations: 8 individual adjustable percentiles from L0,1% to L99,9%, The statistical calculations are performed in real time within each frequency band and for each profile period if period length is set longer than 2 minutes. Class width is 0,2 dB. Dynamic range is 130dB.

Measurement control

Overall measurement duration: 1 sec – 7 days. (If set to Repeat or Synchro mode a new measurement will be started automatically, with no time gap between each measurement).
Time profile A resolution: 5 ms – 24 hours including full frequency spectra. From 1 sec if profile B or Moving is enabled.
Time profile B and Moving resolution: 1 s – 24 hours. A, C and Z – no frequency spectra.
Back erase in Pause mode: 0-20 sec free delectable graphical back erase.

Audio recording

12 and 48 kHz sampling rate / 8 , 16 and 24 bits / 0-120 sec pre-trigger.

Measurement range

One range covering 120dB without any range adjustments.
Self noise measured with 1/2” microphone with a nominal sensitivity of 50mv/Pa: 17 dBA.
Maximum RMS level 137 dBA, Maximum Peak level 140 dB PeakC.

Noise Generator

Built in noise generator with Sine, White, Pink and 1/1 and 1/3 octave band passed filtered noise.

Reverberation Time

Parallel calculation of T15, T20 and T30 in addition to Tmax for Noise excitation and EDT for impulse excitation.

FFT

Sampling rate: 48 kHz.
Frequency resolution: 1.46 Hz or 2.92 Hz (FFT size 8192 or 16384 lines).
Upper frequency: 22 kHz
Functions: Auto spectrum, linear averaging.
Acquisition rate: ~700 ms – no overlap.
Display: Dynamic display zoom 1 – to 64 in power-of-two steps.
Time windowing: Hanning

Display

4,3″ colour display capacitive touch. Protected by an anti-smudge coated and scratch-proof glass.

GPS

Built in GPS
Support for external IP camera. With an external IP camera, you can make event triggered pictures. Pictures captured during a measurement is marked as events in the L/t graph and can be viewed on the Nor145, in NorCloud or  NorReview.

Datastorage / Datatransfer

Storage modes: Manual, Auto, Repeat and Synchro. Synchro synchronise the start of the next measurement to the next full hour.
Internal memory: 350 MB – not used for measurement storage.
Micro SD card: Supports XC and HC standard. No limitation in max. SD card size.
Data transfer via LAN and USB.
Optionally built in GPRS – 3G/UMTS – 4G/LTE and Wifi infrastructure and hotspot.
Internal antenna supports 3G/4G. External antennas with diversity may be connected. External antenna supports also GPRS connection.

Inputs

Microphone input: One microphone input. Supports standard 7 pin Lemo preamplifiers including SysCheck, Microphone heating, TEDS and IEPE.
Polarization voltage: 0
Preamplifier voltage:  ± 15V.
Comment microphone: Via 3,5 pin mini jack.

Analogue outputs

AC out, 100mV full scale on 15 pin I/O socket.
3 pin mini jack headphone socket for replay of voice notes, listen to microphone AC signal or replay of audio recordings. Comment microphone for voice notes is connected to same plug.
Signal generator on 15 pin I/O socket with Pink, White, Bandpass filtered, sine and impulse noise.

Power

Rechargeable 7 Ah Li-Ion batteries with built in fuel gauge.
External power: 9 – 15Vdc 3-5Watt.
Typical battery lifetime: 8 hours.

Dimensons & weight

Dimensions: 235 x 82 x 29 mm (ex. preamplifier and microphone LxWxH)
Weight : 535 g (including preamplifier and microphone)

Accessories

text

Sound level meters

Nor850 Reporting software

Sound level meters

Nor1026 NorReview

Sound level meters

Nor1255 Sound Calibrator

Sound level meters

Nor1256 Sound Calibrator

Sound level meters

NorVirtual App

Sound level meters

NorVirtual

Sound level meters

Nor282 Power amplifier

Sound level meters

Nor277 Tapping machine

Sound level meters

Nor1451 Windscreen, 60mm

Sound level meters

Nor263B Remote hand switch

Sound level meters

Nor1346 Softbag for Nor145

Sound level meters

Nor345 Mains adaptor

Downloads

text
Software

NorConnect

  • NorConnect - 3.0.191212

    NorConnectWindows-3.0.191212.zip (0.08GB)
    Released
  • NorConnect - 5.0.240124

    NorConnectWindows-5.0.240124.zip (0.11GB)
    Released 26/02/2024

    Important notice for Nor850 users!
    Please note that you must upgrade to Nor850 V3.0.210212 or newer together with this version of NorConnect.

    This version of NorConnect is NOT compatible with Nor850 V2!

Software

Nor1026 NorReview

  • NorReview Nor1026 - 5.2.38

    NorReview5.2.38.zip (0.07GB)
    Released 08/03/2019
  • NorReview - 6.3.49

    NorReview-6.3.49.zip (0.1GB)
    Released 22/12/2022
  • NorReview - 6.4.32

    NorReview-6.4.32-1.zip (0.07GB)
    Released 13/03/2024
Software

NorVirtual

  • NorVirtual - 5.0.240212

    NorVirtual-5.0.240212.zip (0.03GB)
    Released 26/02/2024
Software

NorAcal

  • NorAcal v1.0.6 - v1.0.6

    NorAcal_v1.0.6.zip (0.02GB)
    Released 15/03/2018
Instrument software

Instrument software Nor145

  • Chinese fonts -

    chinesefonts.145 (0.02GB)
    Released

    Chinese fonts must be installed separately if needed.

  • Nor145 Service pack - 4.0.10

    SP_4.0.10.145 (0.03GB)
    Released 20/04/2021

    Copy file to root folder on the SD card. See Installing Nor145 version above.

    This service pack is required for Nor145 application 4.0.4707.

  • Nor145 application - 4.0.4707

    App_4.0.4707.145 (0.01GB)
    Released 07/11/2022

    Copy file to root folder on the SD card. See Installing Nor145 version above.

  • Nor145 Service pack - 5.0.14

    SP_5.0.14.145 (0.03GB)
    Released 26/02/2024

    Copy file to root folder on the SD card. See Installing Nor145 version above.

    This service pack is required for Nor145 application 5.0.3812

    NOTE! You have to upgrade your NorConnect software to v. 5.0.240124.

  • Nor145 application - 5.0.3812

    App_5.0.3812.145 (0.02GB)
    Released 26/02/2024

    Copy file to root folder on the SD card. See Installing Nor145 version above.

    Contact distributor regarding price for V5.

  • Installing Nor145 version -

    Note-installing-Nor145-version.txt (0.28KB)
    Released

    NOTE:
    Installing Nor145 version:
    Copy Service pack and application package to the SD card.

    If a new license is needed: Copy License.txt to the same SD card.
    Insert the SD card in your Nor145 instrument and power the instrument.
    Follow the instructions in the installation process.

     

    Please note! Instruments sold in France, Austria, Germany, Spain and Switzerland is subjected to use pattern evaluated software only. Users in these countries are requested to contact their local Norsonic representative to get the current type approved version.

    Achtung! In Deutschland, Österreich, Spanien und der Schweiz unterliegen Schallmessgeräte einer Bauartzulassung / Baumusterprüfung. Die Firmware des Messgerätes ist Bestandteil dieser Bauartzulassung. Führen Sie daher bitte selber kein Firmware-Update durch, sondern wenden sich bei Interesse / Fragen an Ihren lokalen Norsonic-Vertreter!

    Attention ! Les instruments vendus en France, Suisse, Allemagne, Autriche et Espagne doivent être configurés exclusivement avec le logiciel homologué. Dans ces pays, les utilisateurs sont priés de contacter leur représentant local Norsonic pour obtenir la version actuelle approuvée.

    Por favor, tenga en cuenta que los instrumentos vendidos en España, Austria, Alemania y Suiza deben usarse exclusivamente con el software evaluado en el examen de modelo. Se sugiere a los usuarios de estos países que contacten con los distribuidores locales para obtener la versión aprobada en la actualidad.

Software

STIPA IEC 60268-16 ed. 5

  • STIPA - IEC 60268-16 ed. 5 standard -

    (0KB)
    Released

    EC 60268-16 (Ed.5.0 2020): Sound system equipment – Part 16: Objective rating of speech intelligibility by speech transmission index.
    The CIS-value is calculated from the STI-value as specified in: IEC 60849 (Ed. 2.0 1998-02) Sound systems for emergency purposes. (Withdrawn 2008).
    The instrument measures STIPA as specified in Annex B of IEC286-16 by the direct method.

Type approvals

Type approvals Nor145

  • Type approval PTB - Germany -

    PTB-type-approval-Nor145-October-2021.pdf (1464.79KB)
    Released
  • Type approval LNE - France -

    LNE-37403-rev.-0-du-12-fevrier-2021-NOR145.pdf (1595.38KB)
    Released
  • Type approval BEV – Austria - Beilage 1 zu GZ BEV-2022-0.582.186, revision 2

    NOR-145-Beilage-1-02-Anderung.pdf (410.24KB)
    Released 06/07/2023
  • Type approval METAS – Switzerland -

    METAS_Zulassungszertifikat-CH-A3-19108-02_Nor145_Juni-2021.pdf (439.43KB)
    Released
  • Type approval IPQ – Portugal -

    Model_Approval_Nor145_Portugal.pdf (466.35KB)
    Released
  • Type approval CMI – Czech Republic -

    Nor145-Type-approval-Czech.pdf (234.88KB)
    Released
Certificates

Certificates Nor145

  • Declaration of Conformity Nor145 -

    DOC_Nor145-sound-analyser.pdf (36.59KB)
    Released
  • Supplier’s Declaration of Conformity 47 CFR § 2.1077 Compliance Information -

    Suppliers-Declaration-of-Conformity_Nor145.pdf (33.37KB)
    Released
  • UN38.3 Transports of Dangerous Goods Test Report Nor145 -

    UN38.3-test-Report-Shipping_2018-013_Nor145.pdf (923.1KB)
    Released

Support

Do you have any questions
or need support?


Call Icon Mail Icon