The Advanced View graphs for ISO 9614-2 is described below.
Scope dependency

The available Advanced View graphs depend on the current scopes.

FUNC

Press to toggle between different Advanced View graphs.

The first two graphs are available for all scopes.

Multi-function graph

This graph is similar to the Simple View, the only difference being that functions LW, Ieq and Leq are displayed simultaneously.

Dynamic Capability graph

This graph presents the FpI-function (Leq - Ieq) along with the dynamic capability (Ld) obtained from the last probe calibration or verification.

Ld = ∂pI0 - K, where K is either 10 dB or 7 dB for accuracy Engineering or Survey, respectively.

∂pI0 is the FpI measured when the probe is exposed to a sound field which is exactly the same for both microphones.

Delta graph

This graph is only available for the Segment Scope.

∆Ieq: The difference between Ieq in Horizontal and Ieq in Vertical scan. Negative direction value indicates that the direction is opposite in H vs V.

∆Leq: The difference between Leq in Horizontal and Leq in Vertical scan

s: The uncertainty in dB given by the ISO 9614-2 standard.

For Engineering accuracy, ∆Ieq(bi) ≤ s(bi), where bi denotes frequency band i.

Extranous Noise graph

This graph is only available for the Total Scope.

F+/-: The extraneous noise indicator defined by the ISO 9614-2 standard.

For Engineering accuracy, the requirement is F+/- ≤ 3 dB.

For Survey, there is no requirement.

Getting started with Sound Intensity

Intensity theory

Sound Intensity help index