An event can be trigged by the external trigger connected to the I/O port or a level threshold defined in this menu.
These settings controls how the trigger behaves.
- Max Action Time - defines how long the trigger is allowed to last when triggered. If the trigger reaches this time,
the event is stopped.
- Min Event Duration - defines the minimum time the event must be fulfilled. The event action is terminated if the
event is shorter than this time. The functionality is not supported yet.
- Min Duration Out - defines a minimum time the trigger condition must be off for the trigger to stop registering.
- Time Between Events - defines a minimum time before a new event can be triggered after the previous.
- Time span - Here you specify which time span of the day the event trigger settings shall be active.
This is the possible actions for the trigger. If no action is selected, only a marker is registered in the measurement.
- Recording - starts a recording.
- Picture - takes a picture
- Video - starts a video recording
- Digial Output - Sets the digital output according to specification in Digital and Anlog I/O menu.