bg-graphic
Specifications
Specifications

(Data given with 3mA supply current)

The preamplifier has to be supplied by a constant current in the range 3 – 20 mA.
The output voltage will be a fixed DC (about 10 volt) plus the AC-coupled input voltage. Output AC-voltage swing is minimum +/-5 volt if not limited by the voltage for the current source.

Dynamic range: > 120dB
Input impedance: 20 Gohm, 2.5 pF (typical) for Nor1207/BNC 20 Gohm, 18 pF (typical) for Nor1207
Frequency response ± 0.1 dB: (10 Hz – 20 kHz)
Typical attenuation with microphone Nor1228 (18pF) due to capacitive loading: 1 dB for Nor1207/BNC and 6 dB for Nor1207

Maximum SPL with 50mV/Pa microphone:
Nor1207: 136dB
Nor1207/BNC:131dB

Output impedance: 35 ohm typical

Length incl. microphone Nor1228: 96 mm

Diameter:

  • Nor1207/BNC: 14,5 mm

  • Nor1207: 16 mm

Weight incl. microphone Nor1228: 50 g
Power consumption: 30 mW (dependent on supplied current)
EMC classification: As for sound level meter Nor131
Temperature for specified operation: -10°C to +50°C
Temperature for storing: -30°C to +70°C
Humidity for storing and operation: 0 – 90% RH
Microphone heating: No extra heating available

Max length of cable (100pF/m) with a microphone with sensitivity 50mV/Pa and 18pF capacitance (e.g. Nor1228):
Below 137 dB (re. 20 µPa): 30 m
Below 130 dB (re. 20 µPa): 100 m
Below 120 dB (re. 20 µPa): 300 m

Connector type: Two versions available.

  • Nor1207 with TNC connector and

  • Nor1207/BNC with BNC jack.

ICP® is a registered trademark of PCB Piezotronics.

Frequency response for Nor1207:

Nor1207_Freq-resp

Self noise:

Typical equivalent noise (level with a microphone with sensitivity 50mV/Pa and 18pF capacitance re. 20 µPa, e.g. Nor1228): 2 µV A-weighted (12 dB, Max 15 dB); 2,2 µV C-weighted (13 dB).

Noise Nor1207

Applications

What are you going to measure?

Environmental noise

Environmental noise

Reliable noise monitoring with sound level meter, stand-alone terminals or large-scale systems.

Image-107-(1824x2736px)

Building acoustic

Building acoustic is about the sound that travels between rooms - something you'd rather avoid.

Room-acoustics

Room Acoustics

Room acoustics is about the sound quality in a room, for example in a concert hall.

Nor150-sound-intensity-jorge(800x800px)

Sound Intensity

Measure sound level and direction, even in challenging acoustic environments.

Sound-Power

Sound Power

Sound Power is about how much sound energy a source emits per unit of time, measures in watts.

Noise at Work

Noise at Work

Noise measurement at the workplace for easy mapping and documentation of noise levels

Get your quote

 

bg-graphic